계명대학교 의학도서관 Repository

鉛필터의 투과선량을 이용한 15 MV X선의 에너지스펙트럼 결정과 조직선량 비교

Metadata Downloads
Affiliated Author(s)
최태진김진희김옥배
Alternative Author(s)
Choi, Tae JinKim, Jin HeeKim, Ok Bae
Journal Title
의학물리
ISSN
1226-5829
Issued Date
2008
Keyword
AttenuationEnergy spectrumKernel beam convolutionPercent depth dose
Abstract
Recent radiotherapy dose planning system (RTPS) generally adapted the kernel beam using the convolution method for computation of tissue dose. To get a depth and profile dose in a given depth concerened a given photon beam, the energy spectrum was reconstructed from the attenuation dose of transmission of filter through iterative numerical analysis. The experiments were performed with 15 MV X rays (Oncor, Siemens) and ionization chamber (0.125 cc, PTW) for measurements of filter transmitted dose. The energy spectrum of 15MV X-rays was determined from attenuated dose of lead filter transmission from 0.51 cm to 8.04 cm with energy interval 0.25 MeV. In the results, the peak flux revealed at 3.75 MeV and mean energy of 15 MV X rays was 4.639 MeV in this experiments. The results of transmitted dose of lead filter showed within 0.6% in average but maximum 2.5% discrepancy in a 5 cm thickness of lead filter. Since the tissue dose is highly depend on the its energy, the lateral dose are delivered from the lateral spread of energy fluence through flattening filter shape as tangent 0.075 and 0.125 which showed 4.211 MeV and 3.906 MeV. In this experiments, analyzed the energy spectrum has applied to obtain the percent depth dose of RTPS (XiO, Version 4.3.1, CMS). The generated percent depth dose from $6{\times}6cm^2$ of field to $30{\times}30cm^2$ showed very close to that of experimental measurement within 1 % discrepancy in average. The computed dose profile were within 1% discrepancy to measurement in field size $10{\times}10cm$, however, the large field sizes were obtained within 2% uncertainty. The resulting algorithm produced x-ray spectrum that match both quality and quantity with small discrepancy in this experiments.
Alternative Title
Compare the Clinical Tissue Dose Distributions to the Derived from the Energy Spectrum of 15 MV X Rays Linear Accelerator by Using the Transmitted Dose of Lead Filter
Department
Dept. of Medical Biophysics Engineering (의공학과)
Dept. of Radiation Oncology (방사선종양학)
Publisher
School of Medicine
Citation
최태진 et al. (2008). 鉛필터의 투과선량을 이용한 15 MV X선의 에너지스펙트럼 결정과 조직선량 비교. 의학물리, 19(1), 80–88.
Type
Article
ISSN
1226-5829
URI
https://kumel.medlib.dsmc.or.kr/handle/2015.oak/36850
Appears in Collections:
1. School of Medicine (의과대학) > Dept. of Medical Biophysics Engineering (의공학과)
1. School of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학)
공개 및 라이선스
  • 공개 구분공개
파일 목록

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.