계명대학교 의학도서관 Repository

Subsurface calcification of hydrophilic refractive multifocal intraocular lenses with a hydrophobic surface: A case series

Metadata Downloads
Affiliated Author(s)
전종화
Alternative Author(s)
Jun, Jong Hwa
Journal Title
Medicine.
ISSN
1536-5964
Issued Date
2019
Abstract
Rationale:
Opacification of monofocal intraocular lenses (IOLs) of various designs and materials has been reported. Hydrophilic acrylic IOLs are more prone to opacification than hydrophobic IOLs, but IOL surface modification by hydrophobic materials may improve biocompatibility, and few opacifications of such monofocal lenses have been reported to date. However, here we describe the characteristics of opacification of hydrophilic refractive multifocal IOLs with a hydrophobic surface modification in a cluster of patients who underwent uneventful cataract surgery.

Patient concerns:
In this retrospective observational case series, the medical records of 7 patients in whom opacification of the IOL was identified after implantation of LS-313 MF30 (Lentis M plus, Oculentis), from November 2017 to May 2019, were reviewed.

Diagnosis:
All patients had undergone bilateral implantation of LS-313 MF30 IOLs. Ten eyes of 7 patients showed significant opacification at a mean 49.1 ± 10.2 months postoperatively.

Interventions: The IOLs of 4 cases were explanted.

Outcomes:
All of the opacified cases had received LS-313 MF30 IOLs from February 2014 to August 2014 and experienced decreased visual acuity after 44.6 ± 10.5 months. The explanted IOLs of 4 cases were evaluated by scanning electron microscopy (SEM), energy-dispersive X-ray spectroscopy (EDX), Alizarin-red, and von Kossa staining. These explanted IOLs showed fine and evenly distributed, whitish deposits on the entire IOL, particularly below the surface. Although the constituent of the deposits was identified as calcium by Alizarin-red and von Kossa stain, SEM, and EDX analysis showed no surface deposits of calcium. Paraffin-embedded sections of the IOLs were prepared, and calcium deposition was confirmed by EDX analysis at the subsurface region of the IOL.

Lessens:
Significant opacification of these hydrophilic refractive multifocal IOLs with hydrophobic surface modification was found to be due to abnormal calcification of the subsurface of the IOL. Clinicians must be aware of the opacification of this IOL design, despite surface modification. In particular, it should be noted that there is a high likelihood that the patient may experience vision-related symptoms even with moderate opacity and that opacification may lead to a burdensome IOL exchange.
Department
Dept. of Ophthalmology (안과학)
Publisher
School of Medicine (의과대학)
Citation
Seung Pil Bang et al. (2019). Subsurface calcification of hydrophilic refractive multifocal intraocular lenses with a hydrophobic surface: A case series. Medicine., 98(50), e18379–e18379. doi: 10.1097/MD.0000000000018379
Type
Article
ISSN
1536-5964
DOI
10.1097/MD.0000000000018379
URI
https://kumel.medlib.dsmc.or.kr/handle/2015.oak/42644
Appears in Collections:
1. School of Medicine (의과대학) > Dept. of Ophthalmology (안과학)
공개 및 라이선스
  • 공개 구분공개
파일 목록

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.