계명대학교 의학도서관 Repository

Changes in tapered endotracheal tube cuff pressure after changing position to hyperextension of neck: A randomized clinical trial

Metadata Downloads
Affiliated Author(s)
박지훈김지섭
Alternative Author(s)
Park, Ji HoonKim, Ji Seob
Journal Title
Medicine (Baltimore)
ISSN
1536-5964
Issued Date
2021
Keyword
cuff pressureendotracheal tubeextension of necktapered-shaped cuff
Abstract
Background:
Endotracheal tube (ETT) with a tapered-shaped cuff had an improved sealing effect when compared to ETTs with a conventional cylindrical-shaped cuff. Positional change and neck movement typically displace an ETT and change the intracuff pressure. The aim of the current study was to compare the ETT cuff pressure in the TaperGuard ETT vs the conventional ETT after a change from the supine, neutral position to the extension of the neck and semi-Fowler position for thyroid surgery.

Methods:
This prospective, randomized clinical trial included 50 patients undergoing thyroidectomy. Patients were randomly allocated into one of the 2 groups: tracheal intubation with the TaperGuard ETT or with a conventional ETT. The ETT cuff was inflated with air and the ETT cuff pressure was set initially at 20 cmH2O using a calibrated cuff manometer. ETT cuff pressure and distance from carina to ETT tip were measured at supine and semi-Fowler positions with neck extension.

Results:
After the position change, the ETT tip migrated cephalad and cuff pressure increased in the majority of cases. ETT cuff pressure was significantly higher in the TaperGuard group than the conventional group (28.0 ± 6.6 cmH2O and 22.8 ± 4.5 cmH2O, respectively, P = .001). The degree of cephalad displacement of the ETT tip was comparable between the 2 groups (19.4 ± 6.31 mm in TaperGuard group and 21.9 ± 6.9 mm in conventional group, P = .12).

Conclusions:
After the position change from supine to hyperextension of the neck, the ETT cuff pressure was higher in the TaperGuard ETT than in the conventional ETT, although the extent of displacement of the ETT was comparable between the 2 groups.
Department
Dept. of Anesthesiology & Pain Medicine (마취통증의학)
Publisher
School of Medicine (의과대학)
Citation
Park, Ji-Hoon et al. (2021). Changes in tapered endotracheal tube cuff pressure after changing position to hyperextension of neck: A randomized clinical trial. Medicine (Baltimore), 100(29), e26633. doi: 10.1097/MD.0000000000026633
Type
Article
ISSN
1536-5964
DOI
10.1097/MD.0000000000026633
URI
https://kumel.medlib.dsmc.or.kr/handle/2015.oak/43972
Appears in Collections:
1. School of Medicine (의과대학) > Dept. of Anesthesiology & Pain Medicine (마취통증의학)
공개 및 라이선스
  • 공개 구분공개
파일 목록

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.