계명대학교 의학도서관 Repository

IC-10계열의 이온전리함과 TG-21을 이용한 장기간 선형가속기의 출력선량 평가 비교

Metadata Downloads
Author(s)
최태진김진희김옥배
Keimyung Author(s)
Choi, Tae JinKim, Jin HeeKim, Ok Bae
Department
Dept. of Medical Biophysics Engineering (의공학과)
Dept. of Radiation Oncology (방사선종양학)
Journal Title
Keimyung Medical Journal
Issued Date
2011
Volume
30
Issue
1
Keyword
NgasNxOutput doseTG-21TLD monitor
Abstract
This study is to provided the accuracy and stability of the output dose of linear accelerator photon
beams by using the IC-10 and IC-15 air ionization chambers through the Task Group 21 protocol. The gas absorbed dose calibration factor Ngas is determined by using the Bragg-Gray cavity theory for the air exposure calibration factor Nx of Co-60 gamma rays which is delivered from manufacture through NBS calibration. The ionization chamber of IC-10 series composit the Shonka C-552 material with density 1.76 g/ml which is equivalent to air material. the authors
provided the Awall from correction equation for air absorbed calibration factor Ngas and ion
recombination correction factor Aion from Pion analytic measurement. The absorbed dose in water
phantorm is derived from the Ngas including the correction factors mass stopping power ratio of
water to that of gas, ion recombination and replacement for 6 and 23 MV X rays, respectively.
The determined absorbed dose of water phantom exposed to a postal TLD block for output
evaluation in MDACC’s postal monitoring program. The results of comparisons of evaluated absorbed dose to that of MDACC have shown the 1.000 ± 0.012 for 6 MV and 1.001 0.013 discrepancy for 23 MV X rays for a long term followed. This study shows the evaluated outputs for linear accelerator photon beams are very close to that of international output monitor with
small discrepancy of 1.3% for one and half decade.
Alternative Title
The Comparison the Output dose of a Linear Accelerator Photon Beams by Using the Ionization Chamber IC-10 Series Through TG-21 Protocol to Postal Monitoring Output of MDACC for One and Half Decade
Keimyung Author(s)(Kor)
최태진
김진희
김옥배
Publisher
Keimyung University School of Medicine
Citation
최태진 et al. (2011). IC-10계열의 이온전리함과 TG-21을 이용한 장기간 선형가속기의 출력선량 평가 비교. Keimyung Medical Journal, 30(1), 17–28.
Type
Article
URI
https://kumel.medlib.dsmc.or.kr/handle/2015.oak/15515
Appears in Collections:
2. Keimyung Medical Journal (계명의대 학술지) > 2011
1. School of Medicine (의과대학) > Dept. of Medical Biophysics Engineering (의공학과)
1. School of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학)
Authorize & License
  • Authorize공개
Files in This Item:

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.