6 MV 광자선조사면내 투과성필터에 의한 조직선량
- Author(s)
- 최태진; 김옥배
- Keimyung Author(s)
- Kim, Ok Bae; Choi, Tae Jin
- Department
- Dept. of Radiation Oncology (방사선종양학)
Dept. of Medical Biophysics Engineering (의공학과)
- Journal Title
- 의학물리
- Issued Date
- 1993
- Volume
- 4
- Issue
- 1
- Keyword
- Partial Attenuation Filter; TMR; SMR
- Abstract
- Measured and calculated the TMR and SMR factors from percent depth dose underpartial attenuators which cover the whole part of the radiation beam with variousfilter thickness from 0 to 50 mm. This study was performed for x-ray beams generated with a 6 MV linear acceleratorat source to surface distance of 100cm in a water phantom for Lipowitz metal. TMR(0,d,t) was derived from non-linear polynomial regression with field sizedifferencies and a given filter thickness. In this experiments, the TMR(0,10,50) of 50mm of filter thickness was showed13.6 % higher than that of open field and SMR(5,10,50) was 38.5% smaller than thatof open field in same depth.