Abstract
VEP is a gross electrical signal generated by the occipital region of the cortex in response to visual stimuli and it is useful in refraction, optic nerve disease, color blindness, amblyopia, field defect, macular disease, etc.
We used Horizon computer with UTAS-E and normal VEP at different check-size was measured on 30 subjects(60 eyes) with visual acuity better than 1.0
The pattern reversal frequency was 2 Hz., analysis time was 250 msec., artifact reject threshold was 75μV., low-pass filter cut -off was 30Hz.’ high-pass filter cut-off was 1Hz.
The check sizes used were 16x 16(50min), 32x32(25min) and 64x64(12. 5min). The results were as follows;
1. 16x16(50min)
amplitude : 8.18μ3. 55μV.,latency :
2. 32x32(25min) amplitude : 8. 48士5. 99μV., latency
3. 64 x 64(12. 5min)
amplitude : 7.79±3.68μV,latency : 109. 73士5.15 msec.
4. The change of latency between 32 x 32 and 64 x 64 check-ske was statistically significant (p<0.05).
5. The amplitude was largest in 32 x 32(25min.) check-size but statistically not significant (p〉0. 05).
6. The latency was most increased in 64 x 64(12. 5^iin.) check-size and it was statistically significant (p <0. 001).
Alternative Title
VEP in Normal Eyes by Using Different Check-sizes
Keimyung Author(s)(Kor)
김기산
Publisher
School of Medicine
Citation
박태용 and 김기산. (1987). 정상안에서 Check-size에 따른 VEP변화. 대한안과학회지, 28(3), 539–544.