세라믹 골두 골절에서 금속 대 폴리에틸렌 관절면 재치환술 후 발생한 Metallosis - 증례 보고-
- Author(s)
- 이경재; 정수원; 조철현; 민병우
- Keimyung Author(s)
- Lee, Kyung Jae; Cho, Chul Hyun; Min, Byung Woo
- Department
- Dept. of Orthopedic Surgery (정형외과학)
- Journal Title
- 대한정형외과학회지
- Issued Date
- 2008
- Volume
- 43
- Issue
- 4
- Keyword
- Total hip arthroplasty; Ceramic head fracture; Third body wear; Metallosis; 인공 고관절 전치환술; 세라믹 골두 파손; 제 3개체 마모
- Abstract
- A ceramic articulation thought to be superior to metal-on-polyethylene due to its extremely low coefficient
of friction and its potential for resistance to wear. But ceramic is brittle, which makes it theoretically
susceptible to fracture. Although many authors have reported that metal-on-polyethylene articular pairing
had good results after a ceramic head fracture, there have been few reports about complication due to
the remaining ceramic particles after revision surgery. We report here on a case of severe metallosis
that was thought to be due to third body wear by the remaining fractured ceramic particles after revision
total hip arthroplasty.
세라믹은 마찰계수 및 마모에 대한 저항성에 있어 금속-폴리에틸렌 관절보다 우수 하지만 소성 변형 범위가
매우 적어(brittle) 깨지기 쉬운 특성을 가지고 있다. 알루미나 세라믹 골두의 파손은 드물게 보고되고 있으며
골두 파손 시 대퇴 스템의 Morse taper의 소성 변형으로 인해 금속-폴리에틸렌 관절면을 이용한 재치환술을
시행하는 것이 좋은 결과를 보인다고 보고되고 있지만 재치환술 후 남아있는 세라믹 입자에 의한 관절면의
마모에 대한 보고는 매우 드물다. 저자들은 양측성 대퇴골두 무혈성 괴사로 세라믹-세라믹 관절면을 이용한
인공 고관절 전치환술 후 세라믹 골두 골절이 발생한 환자에서 금속-폴리에틸렌 관절면을 이용하여 재치환술을
시행하고 추시 관찰 중 남아있던 세라믹 입자에 의한 제 3개체 마모(third body wear)로 발생한 과도한
metallosis 1예를 경험하였기에 이를 보고하고자 한다.