6MV 광자선의 투과성필터와 Wedge 선속을 이용한 부비강의 균등선량계획
- Author(s)
- 최태진; 이호준; 김옥배
- Keimyung Author(s)
- Choi, Tae Jin; Kim, Ok Bae
- Department
- Dept. of Medical Biophysics Engineering (의공학과)
Dept. of Radiation Oncology (방사선종양학)
- Journal Title
- 대한치료방사선과학회지
- Issued Date
- 1993
- Volume
- 11
- Issue
- 1
- Keyword
- Partial Filter; Homogeneous Dose
- Abstract
- The homogeneous dose planning is one of the most important roles in radiation therapy. But, it is not easy to obtain a homogeneous dose to paranasal sinus region including the ethmoidal sinus with conventional irradiation techniques. In this experimental study, the authors tried to get a homogeneous dose at PNS region, but the nasal cartirage does not exceed the tolerance dose. with anterior-posterior beam and two both lateral wedged beams. Used three fields were shielded with full thickness of blocks to preserve the eye-balls and with blocks of one half value layer to create a homogeneous dose at the whole treatment volume. The dose computations are based on the three dimensonal structure with modified scatter contributions of partial shielders and attenuated beams in 6 MV photon beams. The dose dirtributions of mid-plane is examined with Kodak verification films and teflon-embedded TLD rod (1 mm diameter and 6 mm length) to confirm the computed dose. In our study, the whole PNS regions have shown within 85% of the resultant isodose curves with relatively homogeneous dose distribution. The results of dose computation and measurements are agree well within 5% uncertainties.