Development of a New Radiotherapy Technique using the Quasi-Conformation Method
- Author(s)
- Tae Jin Choi; Jin Hee Kim; Ok Bae Kim
- Keimyung Author(s)
- Kim, Jin Hee; Kim, Ok Bae; Choi, Tae Jin
- Department
- Dept. of Radiation Oncology (방사선종양학)
Dept. of Medical Biophysics Engineering (의공학과)
- Journal Title
- 대한치료방사선과학회지
- Issued Date
- 1991
- Volume
- 9
- Issue
- 2
- Keyword
- Irregular field; Multi-step filter; Quasi-conformation therapy
- Abstract
- The quasi-conformation therapy was performed to get a homogeneous dose distributions for irregeular shaped tumor lesion by using the arc moving beam and beam modifying filer which was made by cerrobend alloy (p=9.4g/cc) metal. In our dose calcuation programme, it was fundmentally based on Clarkson's method to calcuate the irregular multi-step block field in rotation therapy. In this study, the expected relative depth doses under multipartial attenuator agree well with measured data at same plane. The results of comparison the dose computation with that of TLD measurement are very closed within +/-5% uncertainties in the irradiation to phantom with quasi-comformation method. And it has shown that irregular typed multi-step filter can be applied to quasi-conformation therapy in high energy radiation plannings.
- 공개 및 라이선스
-
- 파일 목록
-
Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.