폐동맥 카테터의 하대정맥 캐뉼라 삽입 부위에서의 봉합 -증례보고-
- Author(s)
- 김인수; 이용철; 김진모; In Soo Kim; Yong Cheol Lee; Jin Mo Kim
- Keimyung Author(s)
- Lee, Yong Cheol; Kim, Jin Mo
- Department
- Dept. of Anesthesiology & Pain Medicine (마취통증의학)
- Journal Title
- 대한마취과학회지
- Issued Date
- 2006
- Volume
- 51
- Issue
- 4
- Abstract
- A 63-year-old female with mitral stenoinsufficiency (stenosis and insufficiency) underwent mitral valve replacement. Under general anesthesia, pulmonary artery catheter (PAC) was inserted via the right internal jugular vein. Because the PAC function was normal in perioperative period we didn't recognize the entrapment of the PAC. But on attempting reposition of catheter in postoperative period, we encountered resistance, and postoperative chest X-ray revealed the catheter formed acute angle near right atrial wall. We suspected entrapment of PAC and decided to repeat sternotomy to release the PAC. We found the PAC encircled in the purse-string suture at the inferior vena cava cannulation site. The PAC was removed without any complications. (Korean J Anesthesiol 2006; 51: 486∼9)
Key Words: complication, entrapment, pulmonary artery catheter.