Abstract
A 63-year-old female with mitral stenoinsufficiency (stenosis and insufficiency) underwent mitral valve replacement. Under general anesthesia, pulmonary artery catheter (PAC) was inserted via the right internal jugular vein. Because the PAC function was normal in perioperative period we didn't recognize the entrapment of the PAC. But on attempting reposition of catheter in postoperative period, we encountered resistance, and postoperative chest X-ray revealed the catheter formed acute angle near right atrial wall. We suspected entrapment of PAC and decided to repeat sternotomy to release the PAC. We found the PAC encircled in the purse-string suture at the inferior vena cava cannulation site. The PAC was removed without any complications. (Korean J Anesthesiol 2006; 51: 486∼9)
Key Words: complication, entrapment, pulmonary artery catheter.
Alternative Title
Entrapment of Pulmonary Artery Catheter in the Purse-String Suture at the Inferior Vena Cava Cannulation
Site - A case report -
Keimyung Author(s)(Kor)
이용철
김진모
Publisher
School of Medicine
Citation
김인수 et al. (2006). 폐동맥 카테터의 하대정맥 캐뉼라 삽입 부위에서의 봉합 -증례보고-. 대한마취과학회지, 51(4), 486~489-486~489. doi: 10.4097/kjae.2006.51.4.486
Type
Article
ISSN
0302-5780
DOI
10.4097/kjae.2006.51.4.486
URI
https://kumel.medlib.dsmc.or.kr/handle/2015.oak/40132
Appears in Collections:
1. School of Medicine (의과대학) > Dept. of Anesthesiology & Pain Medicine (마취통증의학)
공개 및 라이선스
  • 공개 구분공개
파일 목록

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.