Evaluation of Two Commercial Kits for Rapid Pathogen Identification Directly From Positive Blood Cultures by Matrix-Associated Laser Desorption/Ionization Time-of-Flight Mass Spectrometry

Metadata Downloads
Author(s)
Sung Gyun ParkDo-Hoon Kim
Keimyung Author(s)
Park, Sung GyunKim, Do Hoon
Department
Dept. of Laboratory Medicine (진단검사의학)
Journal Title
Ann Clin Microbiol
Issued Date
2023
Volume
26
Issue
1
Keyword
Bloodstream infectionDirect identificationMatrix-assisted laser desorption/ ionization time-of-flight mass spectrometry
Abstract
Background:
A bloodstream infection is a life-threatening medical emergency, with a mortality rate of up to 30%. Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) can be used to identify pathogens directly from positive blood cultures. Two commercial preparation kits, SepsiTyper (Bruker Daltonics, Germany) and SepsiPrep (ASTA Corp., Korea), and two MALDI-TOF MS systems, MALDI Biotyper Sirius (Bruker Daltonics, Germany) and VITEK MS PRIME (bioMérieux, France), are available in Korea. We examined these kits and MALDI-TOF MS systems to analyze their performance.

Methods:
We assessed the effectiveness of direct identification using 47 blood cultures and 3 bile cultures positive for microbial growth. The VIRTUO system (bioMérieux, France) was used to incubate the samples after they were collected in Bact/ALERT aerobic and anaerobic bottles. The manufacturers’ protocols were followed for both the SepsiTyper and SepsiPrep kits.

Results:
The SepsiTyper yielded considerably more accurate identifications than did the SepsiPrep, when utilized in MALDI-TOF MS systems (P = 0.0044). However, the SepsiPrep was easier to use and the results more quickly obtained than with the SepsiTyper. The MALDI Biotyper Sirius produced more accurate identifications with the SepsiTyper than did the VITEK MS PRIME (P = 0.0736). The SepsiTyper enabled the accurate identification of five of six polymicrobial cases, utilizing either the MALDI-TOF MS systems.

Cnclusions:
Among the pathogen ID kits tested in this study, the SepsiTyper with MALDI Biotyper Sirius performed the best. In clinical laboratories utilizing VITEK MS PRIME, it is recommended that the either the SepsiTyper or SepsiPrep kit be used for direct identification, while considering certain limitations in terms of performance.
Keimyung Author(s)(Kor)
박성균
김도훈
Publisher
School of Medicine (의과대학)
Type
Article
ISSN
2288-6850
Source
https://www.acm.or.kr/article_aws/?num=ACM_23-004
DOI
10.5145/ACM.2023.26.1.3