계명대학교 의학도서관 Repository

Comparison of Thick Biolimus A9-Eluting Stent and Thin Zotarolimus-Eluting Stent in Multi-Vessel Percutaneous Coronary Intervention

Metadata Downloads
Author(s)
Cheol Hyun LeeHee Jeong LeeTae-Wan ChungSeonhwa LeeJongmin HwangIn-Cheol Kim
Keimyung Author(s)
Lee, Cheol HyunLee, Hee JeongChung, Tae WanLee, Seon hwaHwang, Jong Min
Department
Dept. of Internal Medicine (내과학)
Journal Title
Korean Circ J
Issued Date
2025
Volume
55
Issue
5
Keyword
Drug-eluting stentPercutaneous coronary interventionPolymer
Abstract
Background and Objectives:
There are limited randomized studies on patients undergoing multi-vessel percutaneous coronary intervention (PCI) comparing the outcomes between stent thickness and polymer types. To compare the clinical outcomes of thick biodegradable polymer-based biolimus A9-eluting stents (BESs) and thin durable polymer-based zotarolimus-eluting stents (ZESs) in patients undergoing multi-vessel PCI.

Methods:
A total of 936 patients who underwent multi-vessel coronary artery stenting were randomly assigned to the BES (n=472) or ZES (n=464) groups. The primary endpoint was 2-year major adverse cardiac events (MACEs), a composite of all-cause death, myocardial infarction (MI), and any revascularization at the 2-year follow-up.

Results:
Fifty-two (11.2%) of 472 patients in BES group and 50 (10.9%) of the 464 patients in ZES group met the 2-year primary endpoint of MACE (hazard ratio, 1.00; 90% confidence interval, 0.72, 1.38; p=0.994). All-cause death (BES vs. ZES: 2.8% vs. 2.7%, p=0.758), MI (2.1% vs. 2.6%, p=0.483), and repeat revascularization (6.7% vs. 6.9%, p=0.876) were not significantly different between the 2 groups. Although there was no significant outcome difference in any subgroup analysis, the technical failure rate leading to the use of other stents was higher in BES than in ZES (3.2% vs. 0.9%, p=0.023).

Conclusions:
In patients who underwent multi-vessel PCI, BES and ZES showed comparable 2-year clinical outcomes. However, BES was not established to demonstrate non-inferiority to ZES in terms of the incidence of the primary endpoint at the 2-year. The technical success rate of the index PCI with the assigned stent was higher for thinner ZES.
Keimyung Author(s)(Kor)
이철현
이희정
정태완
이선화
황종민
김인철
Publisher
School of Medicine (의과대학)
Type
Article
ISSN
1738-5555
Source
https://www.e-kcj.org/DOIx.php?id=10.4070/kcj.2024.0101
DOI
10.4070/kcj.2024.0101
URI
https://kumel.medlib.dsmc.or.kr/handle/2015.oak/46130
Appears in Collections:
1. School of Medicine (의과대학) > Dept. of Internal Medicine (내과학)
공개 및 라이선스
  • 공개 구분공개
파일 목록

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.